Regurgitant Flow Rate
Mitral Regurgitation
· Dr. Kenneth Gross ·
UNMC Advertisement
  Decrease Font Size  Default Font Size  Increase Font Size
Log In ·  Subscribe
Calcs>Mitral Regurgitation>RFR>1
Calculator
Radius
cm
VA cm/sec
Angle degrees
Heart Rate bpm
 
RFR ml/min

Regurgitant Flow Rate

 
How to get a Regurgitant Flow Rate in Mitral Regurgitation.
 
Step 1:  Obtain a CFD of the MR Jet with a good PISA formation under zoomed view.  The midesophageal view with a good quality, largest PISA formation should be used.
 
Step 2: Measure the Diameter of the PISA radius. The radius is from the center of the PISA formation, typically from the center of the valve, to the first aliasing velocity.  If the PISA is small, try slightly decreasing the aliasing velocity (VA) which will increase the size of the PISA formation.
 
Step 3: Obtain the angle of the PISA formation. The angle is typically 180 degrees.
 

Previous Page   Next Page